11.13.3 Messages of Ventilator Alerts
In the event of one or more alerts related to the ventilator, the following messages may be displayed:
Table 25 - Warning Messages
Message
Delay time
Description
ASSISTED FLOW
TRIGGER
< 1
It indicates the occurrence of a trigger
assisted generated by increasing inspiratory
flow.
ASSISTED PRESSURE
TRIGGER
< 1 s
It indicates the occurrence of a trigger
assisted generated by a pressure drop.
SPONTANEOUS FLOW
TRIGGER
< 1 s
It indicates the occurrence of a spontaneous
trigger generated by increasing inspiratory
flow.
SPONTANEOUS
PRESSURE TRIGGER
< 1 s
It indicates the occurrence of a spontaneous
trigger generated by a pressure drop.
MANUAL TRIGGER
(RED)
< 1 s
It indicates the occurrence of a trigger
assisted, manually generated by the
operator.
MANUAL TRIGGER
(YELLOW)
< 1 s
It indicates the occurrence of a spontaneous
trigger generated manually by the operator.
INVERSE I:E RATIO
< 1 s
It indicates that ratio I:E It is inverse, i.e. the
time of the inspiratory phase is greater than
the time of the expiratory phase.