62
MAGNAMED TECNOLOGIA MÉDICA S/A
The internal flow transducer contains two sensors, a flow sensor and the other to measure the
temperature.
Each sensor has a nonlinear output voltage independent. To determine the mass flow of gas
passing through the transducer, the output voltage of each sensor is measured.
A microprocessor processes the results and calculates the flow using a specific algorithm.
The circuit measures the flow is generally known as a thermal sensor or hot wire anemometer.
This flow transducer uses a heated wire sensor and maintained at a temperature of 150 ° C.
The gas velocity through the sensor determines the heat transfer rate between the gas and
sensor.
This heat transfer rate is translated into a voltage required to maintain the temperature at 150 °
C.
Therefore, this voltage is a function of mass flow of gas through the sensor.
The heat transfer rate is also influenced by gas temperature.
A thermistor circuit is used to measure the gas temperature and a correction is made also
through specific algorithm.
Reading Range
Air: 0 to 300 SLPM
O
2
: 0 to 300 SLPM
Specified Tolerance
Air: 2.0% or 0.05
SLPM (Whichever is greater)
O
2
: 2.0% or 0.05
SLPM (Whichever is greater)
Resistance
< 2.5mbar
Gas temperature range
5 to 46°C
Humidity Range
Dry gas (< 10% UR)
Operating Pressure
Atmospheric Pressure
Power Supply
5V ±10% sensor and 2.7V – 5.5V Eeprom
Answering time
< 2.5ms
Rupture Pressure
Rate above 100 psi
Weight
21g
Distal Flow Sensor Specifications
Table 17
– Flow sensor distal – General Specifications
General Specifications
Содержание Fleximag
Страница 1: ...Operation Manual Mechanical Ventilator Fleximag display 15...
Страница 24: ...23 MAGNAMED TECNOLOGIA M DICA S A OK Item Assembly Sequence Image 8 Connect the power cord to the equipment...
Страница 84: ...83 MAGNAMED TECNOLOGIA M DICA S A DualPAP Dual Continuous Positive Airway Pressure...