53
MAGNAMED TECNOLOGIA MÉDICA S/A
measurement is consumable, and the life of the cell, according specification of the original manufacturer is
10,000 hours to 100% O2, i.e. more than one year of continuous use. However, we recommend trading in
maintenance preventive in 24 months schedule or 10,000 hours (whichever occurs first).
Cell Paramagnetic - generates an electrical signal proportional to the concentration of oxygen in the mixture
gas administered to the patient and the intensity of this electrical signal is due to the torque created in the
arrangement Magnetic cell. This sensor utilizes the paramagnetic susceptibility of the oxygen that distinguishes
it from other gases. This method of measurement is not consumable.
•
The galvanic cell for measurement of oxygen concentration should be replaced as indicated in
chapter 11.14.
•
His replacement should be performed by trained and qualified personnel and their disposal must
follow the local regulations.