103
MAGNAMED TECNOLOGIA MÉDICA S/A
11.17.4 Electrical safety
The following are the precautions that should be observed when combining these items (not medical
equipment) with the system.
•
Items that do not comply the requirements of IEC 60601-1 should not be placed within 1.5 m of
the patient
.
•
All items (electromedical equipment or not) connected to the system output signal cable / input
must receive power from a source of alternating current using separate transformer (according
to IEC 60989) or have protective conductor additional land.
•
Do not connect directly electrical equipment nonmedical the AC wall outlet. Using AC power
supply with own transformer. Otherwise, the leakage current increase above the levels accepted
by the IEC 60601-1 under normal terms and conditions of a single failure. This may cause
dangerous electrical shock to the patient or operator.
•
After connecting any equipment into these outlets, subject the system to a complete test leakage
current (according to IEC 60601-1).
•
The electromedical system operator shall not touch electrical equipment no doctor and the
patient simultaneously. This may cause dangerous electrical shock to the patient or operator.