76
MAGNAMED TECNOLOGIA MÉDICA S/A
11.10.6 P-SIMV
P-SIMV
–
Pressure Synchronized Intermittent Mandatory Ventilation
Description:
P-SIMV, secure the respiratory rate,
inspiratory pressure and inspiratory
time, beyond the sensitivity criteria for
the occurrence of ventilator shooting by
the patient.
This mode allows the ventilator to apply
the mandatory cycles predetermined in
sync with the inspiratory effort of the
patient.
The mandatory cycles occur in the
predetermined time window (according
to the set respiratory rate), but
synchronized with the patient trigger.
If there is an apnea, the next cycle will
be triggered by time until they return
the inspiratory incursions patient
To obtain this IMV mode, just off the
support pressure support setting the
zero pressure (ΔPS = 0) or the flow and
pressure sensitivities equal to zero.
Set parameters:
•
PRESSURE INSPIRATORY
•
FREQUENCY
•
TIME INSPIRATORY
•
PEEP
•
CONCENTRATION
•
RISE TIME
•
ΔPS (Pressure Support
- PEEP)
•
FLOW CYCLING (% FLOW)
•
TRIGGER PRESSURE
•
TRIGGER FLOW
•
FLOW (
- NEONATAL)
Note: automatic Backup
(1)
1
–
Whenever the apnea set time is reached, the ventilator
triggers a ventilatory cycle whose configuration is based on the
current mode settings.