60
•
Adjustment of backup ventilation parameters (backup) is only available in spontaneous
ventilatory modes. In the other modes, the backup ventilation is automatic and considers the
parameters adjusted for the ventilatory mode itself.
•
When the operator sets pressure or flow trigger, the controlled modes (VCV, PCV, PRVC and
PLV) will be monitored. In this case, such information will be displayed on the active mode
button.
After adjusting the parameters, for them to be activated, it is necessary to press the CONFIRM button.
To cancel the settings and to continue with the previous settings, including the ventilatory mode, simply
press the CANCEL button. This way the ventilator will ignore the settings made on this screen and will return
to the main screen.
•
If the new value is not confirmed, it will be discarded after the screen lock time.
6.4 Calibrations
It is recommended to calibrate some components before use in the following situations:
6.4.1 Distal Flow Sensor
•
Replacement of the distal flow sensor;
•
Active alarm with message "CHECK FLOW SENSOR".
6.4.2 Integrated valve
•
Replacement of integrated valve;
•
Replacing the diaphragm;
•
Incorrect control of PEEP;
•
Excessive leakage.
6.4.3 O
2
cell (galvanic cell only)
•
Cell replacement;