45
•
Never start the autotest procedure with the ventilator connected to the patient.
If you heard the speaker test, answer YES. Otherwise, answer NO. In this case the equipment shall
not be used and therefore, it will be inoperative until a new autotest is carried out, that is, until the equipment
is restarted.
To start the autotest, press the confirm button or, if you want to start the ventilation immediately, press
the ventilation button.
•
Never respond YES if you could not hear the speaker test under the risk of
malfunctioning of the priority alarms during ventilation.
•
Always perform the autotest procedure before connecting the ventilator to a patient.
The autotest checks all important items for proper ventilation:
•
Gas regulated pressure
•
Proportional Valves
•
Flow Sensors
•
Expiratory valve
•
O
2
cell
•
Leakage
•
Resistance and compliance of the respiratory circuit
•
Flowair system
1
At the end of the leak test, the RESPIRATORY CIRCUIT OUTPUT SHALL BE UNOBSTRUCTED to
perform the resistance test and, where appropriate, the proximal flow sensor test.
•
Do not forget to unobstructed the circuit output before the resistance test.
1
Only for Fleximag Max 300 and Fleximag Max 700