Nebulization or humidification can increase the resistance of respiratory system filters.
The operator shall often monitor increased resistance and filter blockage of the
respiratory system.
•
The ventilator cannot be used in a hyperbaric chamber.
•
Only use air and oxygen in the gas inlet of the equipment. Do not use nitric oxide,
helium or other gases.
•
The expiratory branch may become contaminated with bodily fluids or expired gases
during use of the equipment under normal conditions and under a single fault.
•
HME filter, HEPA filter and airway adapter are single use. The reuse of these
accessories may cause cross contamination.
•
The distal flow sensor is for single patient use and shall be discarded when changing
patients. Re-use for another patient may result in cross-contamination.
•
FlexiMag Max does not generate sub atmospheric pressure during the expiratory phase.
1.3 Attention
•
This ICU ventilator does not emit electromagnetic waves that may interfere with the
operation of equipment in its vicinity.
•
Perform all maintenance according to the operating time specified in this manual.
•
All service, modification or maintenance on the ventilator can only be carried out by
qualified technician, trained and duly authorized by MAGNAMED.
•
Only use parts, cables, sensors, filters and breathing circuits specified by MAGNAMED
and compatible with the system. For purchase, please inform the codes presented in
the corresponding chapter.
1.4 Observation
•
The technical specifications of MAGNAMED products are subject to change without notice.