![Magnamed Fleximag Max 300 Скачать руководство пользователя страница 147](http://html1.mh-extra.com/html/magnamed/fleximag-max-300/fleximag-max-300_operation-manual_691599147.webp)
147
•
Use only MASKS specified by MAGNAMED with local registration and CE mark.
•
Use MASK suitable for patient type to avoid excessive leakage.
10.27 Respiratory Circuit
Table 56
– Respiratory circuit specification
Specification
Adult/ pediatric connection
22 mm
Neonatal connection
15 mm
Resistance
≤ 0.3 mbar/L.s-1
•
Use only respiratory circuit specified by MAGNAMED with local registration and CE mark.
•
Use respiratory circuit suitable for patient type to avoid excessive leakage.
10.28 HME Filter
•
It is recommended the use of HME filter following the specifications given below with
local registration and CE mark.
Table 57
– HME filter specification
Specification
Compliance
ISO 23328-1, ISO 23328-2, ISO 9360-1 and
ISO 9360-2