137
Table 43
- Operation Specifications
Description
Specification
Tolerance
Recommended Maximum Operating
Temperature
37 °C
± 3 °C
Recommended maximum operating
pressure
120 cmH2O
± 2 cmH2O
10.21 Maintenance and Calibration Specifications
Table 44 - Maintenance and calibration specifications
Description
Specification
Tolerance
Inspection and replacement of
integrated valve diaphragm
Under inspection or 5,000 hours or 12 months
(whichever comes first)
500 h /
1 month
Inspection and replacement of the O
2
galvanic cell
Recommended replacement if there are problems with
calibration or 10,000 hours or 24 months (whichever
comes first)
inspection and Replacement of
Internal Battery
10,000 hours or 24 months (whichever comes first)
Equipment Inspection
5,000 hours or 12 months (whichever comes first)
Calibration of equipment
5,000 hours or 12 months (whichever comes first)
10.22 IRMA CO
2
Sensor Specifications
Table 45 - IRMA CO
2
sensor - General specifications
General specifications
Description
Mainstream monitoring sensor with infrared technology.
Dimensions (L x P X A)
38 x 37 x 34mm (1.49” x 1.45” x 1.34”)
Cable length
2.50m (± 0.02m)
Weight
<25g (without cable) | <38g (with cable)
Atmospheric Operating Pressure
525 to 1200cmH2O (525cmH2O corresponds to an altitude of
4572m or 15,000 feet).
Storage and Transport Atmospheric
Pressure
500 to 1200cmH
2
O.
Mechanical resistance
Supports 1m repeated drops on a hard surface.
Power Supply
4.5 to 5.5 VDC, Max 1.0W (power measured with 5V and less
than 350mA for 200ms).
Surface Temperature (Ambient
Temperature of 23 ° C)
Max: 41°C / 106°F.
Airway Adapter
Adult / Pediatric (Disposable):
Adds less than 6ml of dead space;