119
Item
Parameter
Specification
Resolution
Unit
27
Volume of the sigh
(6)
10 to 100
10
% Vt
28
Rate of the sigh
(6)
20 to 100
10
cycles
29
Tube compensation
Endotracheal
Tracheostomy
---
---
30
Diameter of the tube
2.5 to 12.0
2.5 to 10 : 0.5
mm
10 to 12 : 1
31
% Tube Compensation
10 to 100
10
%
32
Minimum inspiratory hold
(7)
Minimum expiratory hold
0,1 to 30
0,1 to 1 : 0.1
s
1 to 30 : 1
33
Alarm muting time
OFF, 10 to 120
10
s
34
Screen Lock Time
OFF, 1 to 30
1 to 5: 1
min
10 to 30: 5
35
Flow (only in O
2
Therapy
mode)
0 to 60
0 to 60: 1
L/min
(1) Neonatal patient volume only with NEONATAL proximal flow sensor and in VG mode.
(2) Respiratory rate 0 (zero) will only be reached in spontaneous modes, with triggers and backup
off.
(3) The minimum and maximum values of rate and I: E ratio depend on the adjusted ventilatory
mode.
(4) Adjustable backup options for spontaneous modes, for other modes, backup is automatic.
(5) Nebulizer and TGI flows cannot be activated simultaneously.
(6) Sigh adjustment available only in VCV and V-SIMV modes
(7) Hold time when immediately pressing and releasing button.
•
The FlexiMag Max ventilator cares for any patient, from premature to morbidly obese,
however, the patient height adjustment used for optimal weight calculation is limited.
•
For patients who exceed this limit, the parameters can be adjusted directly by the
operator.