background image

500 D 01/2012

17a

3.5   Special functions

3.5.1 Display of extinction

After measurement 

17.11.11  16:06
0042 AMMONIUM 10   

0001 place 

8,0 mg/L 

NH4

1+0   

   

the extinction of the measured value can be displayed by pressing  .

17.11.11  16:06
0042 AMMONIUM 10   

0001 place 

1+0   

1.090 E

Display of extinction, when the measuring range is exceeded

After measurement 

16.11.11  16:06
0042 AMMONIUM 10   

0001 place 

>10,0 mg/L 

NH4

1+0   

the extinction of the measured value can be displayed by pressing  . When  

releasing key  , the calculated measuring value is displayed. This estimated value 

is not stored; in the above example > 10,0 mg/L NH4 will be stored.

17.11.11  16:06
0042 AMMONIUM 10   

0001 place 

1+0   

2.068 E

   

17.11.11  16:06
0042 AMMONIUM 10   

0001 place 

15,1 mg/L 

NH4

1+0   

3.5.2 Reaction time

If you wish, the reaction time can be displayed before measurement. Press key 

OK

 

to start the clock.

17.11.11  16:06
0042 AMMONIUM 10   
React.time 

15’00

The reaction time can be skipped or stopped at any time using key 

U

 or 

S

After completion of the reaction time the sample is measured automatically, if there 

is a cuvette in the cuvette slot. The configuration menu allows to completely activate 

or deactivate the reaction timer.

Содержание NANOCOLOR 500 D

Страница 1: ...ax 41 62 388 55 05 E Mail sales ch mn net com Deutschland und weltweit Germany and worldwide Tel 49 24 21 969 0 Fax 49 24 21 969 199 E Mail info mn net com www mn net com Bedienungsanleitung Manual NA...

Страница 2: ...CS 15a 3 4 5 905 TRANSMISSION 16a 3 5 Special functions 17a 3 5 1 Display of extinction 17a 3 5 2 Reaction time 17a 3 5 3 Memory administration 18a 3 5 4 Special user defined methods 21a 3 5 5 Numberi...

Страница 3: ...splay etc 1 1 Technical data Technical data Single beam filter photometer for a wavelength range from 340 to 860 nm Silicon photo element Automatic zero adjustment Graphic display 128 x 64 pixels Foil...

Страница 4: ...ner 4 cuvette slot with automatic cuvette identification 5 socket for connection of mains adaptor 6 RESET push button subsurface 7 USB 1 1 interface for data transfer and update 8 RS 232 C interface f...

Страница 5: ...olored glass filter Other interference filters on request 1 4 Power supply and light source Rechargeable battery The built in rechargeable battery allows stationary and mobile operation of the pho tom...

Страница 6: ...election of a method via method number is always pos sible With the photometer NANOCOLOR 500 D the barcode on the test tubes performs this feature thus simplifying the measurement and minimising possi...

Страница 7: ...1 NANOCOLOR 5OOD NANOCOLOR 5OOD please wait 2 After this self check the method selection mode is activated 17 11 11 16 06 cuvette Method ___ 3 If during the self check the photometer measures a deviat...

Страница 8: ...on Key U Roll mode upward key interrupt of the automatic program sequence method selec tion call up previous submethod next program step previ ous sampling location etc Key U Roll mode downward key in...

Страница 9: ...calibra tion measuring direction Method call up Many analytical parameters can be measured with different cuvette sizes measur ing ranges dimensions and measuring wavelengths The method can be called...

Страница 10: ...g key F2 after a measurement as long as the cuvette is still in the cuvette slot you can assign a 12 character alphanumeric name for the sampling place After input this name is displayed in the lower...

Страница 11: ...filter is automatically turned into measuring position and the cuvette is measured immediately 17 11 11 16 06 0261 COD 160 0001 place 22 mg L O2 1 0 4 If desired place other sample test tubes in phot...

Страница 12: ...16 06 0042 AMMONIUM 10 React time 15 00 6 Place test tube with sample in photometer start reaction time with key OK or skip with key U The test tube with base line value NULL is not used 17 11 11 16...

Страница 13: ...1 99_ 17 11 11 16 06 0042 AMMONIUM 10 1234 sewage B 370 0 mg L NH4 1 99 10 If necessary place other test samples in photometer and measure samples are numbered in sequence Current sample number place...

Страница 14: ...tus line between time and battery indicator automatic storage of results is activated see Configuration chap 3 5 3 This means that the respective result is automatically stored when the cuvette is rem...

Страница 15: ...ions tube tests standard tests BIO tests 17 11 11 16 06 Method 901 FACTOR 902 STANDARD 903 EXTINCTION 904 KINETICS 3 4 1 901 FACTOR Since the color reactions normally used for measurements obey the La...

Страница 16: ...medications In such cases it is necessary to analyze a standard with a known concentration as well as the test sample The test results are then related to those of the standard The Lambert Beer Law E...

Страница 17: ...d wavelength using keys U U and confirm the selection by pressing OK The selected wavelength is adjusted 3 Perform measurement following the instruction of the display 4 Return to method selection usi...

Страница 18: ...ough test sample x 100 light transmited through blank value The dimensionless variable light transmited through test sample is referred as the transmittance light transmited through blank value Measur...

Страница 19: ...the calculated measuring value is displayed This estimated value is not stored in the above example 10 0 mg L NH4 will be stored 17 11 11 16 06 0042 AMMONIUM 10 0001 place 1 0 2 068 E 17 11 11 16 06...

Страница 20: ...ata sets You can either accept this message by pressing OK and continue measurements without storage or you can call up the memory administration by pressing key MEM transfer the con tents of the memo...

Страница 21: ...ey OK at this point you can display all data sets in sequence us ing key U If you have previously made a selection only the selected data sets are displayed 17 11 11 16 06 17 11 11 16 06 MEMORY 0761 t...

Страница 22: ...s characters the instrument selects all names starting with this character string In this case it does not matter whether you enter upper case letters or lower case letters Here again you can skip the...

Страница 23: ...supplement and reaction time every letter or number can be entered directly via the keyboard When the display asks factor linear you can either press key OK and enter a linear factor E x F where the...

Страница 24: ...on with OK all previous entries can be edited 17 11 11 16 06 P56 Selenium edit Output Using keys U U you access the submenu output After confirmation with OK all previous entries can be transferred to...

Страница 25: ...e for a given sample number Generally numbering of samples starts with 0001 after every method selection 1671 NITRITE 0001 place 1 0 sample No _ 17 11 11 16 06 1671 NITRITE 0001 place 1 0 sample No 12...

Страница 26: ...ether with all other parameters of the current data set and the resulting measuring value is directly calculated and displayed When you remove the cuvette the complete data set is stored 1671 NITRITE...

Страница 27: ...and confirm with OK 17 11 11 16 06 CONFIGURATION GB date time calibrate memory date 17 11 11 16 06 time CONFIGURATION With keys U U select date or time and adjust using the keyboard 3 Select calibrati...

Страница 28: ...and confirm with OK 17 11 11 16 06 CONFIGURATION COM 1 calibrate memory react time COM 1 PC 19200 printer 9600 17 11 11 16 06 CONFIGURATION COM 1 ASCII SDF SDF 17 11 11 16 06 CONFIGURATION Adjust the...

Страница 29: ...t AUTO OFF and confirm with OK 17 11 11 16 06 CONFIGURATION AUTO OFF signal tone heading filter 11 12 CONFIGURATION AUTO OFF 6O min 17 11 11 16 06 Using keys U U you can determine the turn off time of...

Страница 30: ...the photometer has to be connected to the computer either via the serial or the USB interface Start the update pro gramme on the PC and on the photometer After a successful update the pho tometer ret...

Страница 31: ...Hyperterminal For serial transfer it is important to match the transfer rate between photometer and computer Basic setting 19200 baud 8 data bits no parity 1 stop bit no protocol The configuration ha...

Страница 32: ...ound during self check or cali bration restart calibration from the con figuration menu If error remains contact your local distributor for filter replacement lamp defective no light at the photo cell...

Страница 33: ...NANOCOLOR NanOx at elevated pressures in a microwave oven Very resistant samples applicable to points 1 2 3 and 7 above can be treated by wet decomposition oxidation with nitric and sulfuric acid Inst...

Страница 34: ...ding to the following table Method Lead 5 test 0 09 0 10 5 00 mg L Pb 0093 Cadmium 2 test 0 14 0 10 2 00 mg L Cd 0143 Iron 3 test 0 37 0 10 3 00 mg L Fe 0373 Copper 7 test 0 54 0 10 7 00 mg L Cu 0543...

Страница 35: ...nt set NANOCOLOR NanOx Metal REF 918 978 Requisite accessories NANOCOLOR heating block test tubes 14 mm ID REF 916 80 or alternative ly microwave oven PTFE pressure vessel 20 mL for decomposition and...

Страница 36: ...mg L Cd 0142 Chromate 5 test 0 24 0 05 2 00 mg L Cr 0244 Chrome 0 005 0 50 mg L Cr 1244 Iron 3 test 0 37 0 10 3 00 mg L Fe 0372 Copper 7 test 0 54 0 10 7 00 mg L Cu 0542 Nickel 7 test 0 61 0 10 7 00...

Страница 37: ...ion in a heating block at 100 120 C features a lower oxidation potential than microwave decomposition For mainly municipal waste waters this method can be used if the matrix is constant over a longer...

Страница 38: ...account Only the main reagent which forms the measured color complex is not added With the NANOCOLOR photometer the measurement programme for the correc tion value is started after the measurement of...

Страница 39: ...26 0 27 0 28 0 29 0 30 0 33 COD 40 60000 Almost all colors and turbidities are destroyed under test conditions and do not interfere COD resistant colors and turbidities cause deviating results which c...

Страница 40: ...mL distilled water close mix 0 79 ortho and total Phosphate 50 Proceed as described in the instructions for test 0 79 but instead of R3 add 1 0 mL sulfuric acid 20 close mix 0 82 Oxygen 12 8 22 8 25...

Страница 41: ...sample 2 mL R1 mix fill to 25 mL with distilled water mix 1 42 Fluoride no correction possible 1 44 Hydrazine 20 mL sample 2 mL R1 mix fill to 25 mL with distilled water mix 1 483 Silica high sensitiv...

Страница 42: ...mix 20 mL sample 1 mL R1 mix 5 drops of R3 mix fill to 25 mL with distilled water mix 1 25 Chromate 2 0 mL R2 20 mL sample mix fill to 25 mL with distilled water mix 1 30 Cyanide 20 mL sample 1 spoon...

Отзывы: