500 D 01/2012
25a
3.6
Configuration of the
NANOCOLOR
®
500 D
In method selection mode enter
) * *
. All topics of the configuration can be
selected in the scroll mode using keys
U
/
U
.
1. Select language and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
(GB) ---> (?)
date / time
calibrate
memory
→
17.11.11 16:06
CONFIGURATION
language: (GB) ?
Wait until your language is displayed, then press key
OK
until confirmation
appears. You can choose between deutsch (D), english (GB), français (F), ital
-
iano (I), nederlands (NL), español (E), magyar (H), polski (PL), português (P),
esky (CZ), baha indonesia (INA) and slovensko (SLO).
2. Select date / time and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
(GB) ---> (?)
date / time
calibrate
memory
→
↑
date
17.11.11 16:06
↓
time
CONFIGURATION
With keys
U
/
U
select date or time and adjust using the keyboard.
3. Select calibration and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
(GB) ---> (?)
date / time
calibrate
memory
→
17.11.11 16:06
CONFIGURATION
calibrate
back
→
CALIBRATION
17.11.11 16:06
please wait
First the photometer automatically checks all filters and the lamp. For the second
calibration step the
NANOCOLOR
®
500 D requires a calibration tube (a clean
NANOCOLOR
®
500 D test tube filled with water).
CALIBRATION
17.11.11 16:06
insert cal.-tube
CALIBRATION
17.11.11 16:06
please wait
The calibration values are automatically measured, stored and used for all
following measurements
17.11.11 16:06
please remove
cuvette