500 D 01/2012
27a
8. Select heading and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
COM 1
signal tone
heading
react.time
→
17.11.11 16:06
CONFIGURATION
NANOCOLOR 5OOD_
Using keys
U
/
U
you can edit the heading or you may enter a new heading
using the keyboard of the photometer.
9. Select special filters 11 and 12 and confirm with
OK
.
17.11.11 16:06
CONFIGURATION:
COM 1
signal tone
heading
filter 11/12
→
17.11.11 16:06
CONFIGURATION
filter 11 ...nm
filter 12 ...nm
↑
↓
Using keys
U
/
U
select position 11 or 12. Enter the wavelength(s) of the
additional special filter(s) via the keyboard.
10. Select AUTO-OFF and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
AUTO-OFF
signal tone
heading
filter 11/12
→
CONFIGURATION
AUTO-OFF
6O min
17.11.11 16:06
↑
↓
Using keys
U
/
U
you can determine the turn-off time of the photometer after
the last measurement. Available times run from 10 to 120 min in 10-minute inter-
vals (ex factory = 20 min). Confirm your setting with
OK
.
11. Select SYSTEM-RESET and confirm with
OK
.
17.11.11 16:06
CONFIGURATION
AUTO-OFF
SYSTEM-RESET
heading
filter 11/12
→
CONFIGURATION
SYSTEM-RESET
back
17.11.11 16:06
→
After a second confirmation the photometer performs a SYSTEM-RESET, thus
restoring the state of delivery, i. e., the complete memory, all user-defined spe
-
cial methods etc. are delete. The display then asks for a calibration. Confirm with
OK
and perform the calibration.