39a
500 D 01/2012
5.11.2 Determination of correction values for NANOCOLOR
®
standard tests with
reagent blank value
For tests of this category, the instructions require distilled water plus reagents as a
reagent blank value. Another volumetric flask 25 mL is necessary for the correction
value. Distilled water is used as a blank value for the correction value.
For some tests, a special reagent has to be prepared for correction. If reduced
sample volumes are used for a test (dilution due to concentrations which are too
high), the sample volume for determination of the correction value must be reduced
by the same amount.
Test
Preparation for correction (value B)
1-02 Aluminium
20 mL sample; 0.2 mL R1, mix; 1 measuring spoon R2, mix;
2 mL R4, mix; fill to 25 mL with distilled water, mix.
1-05 Ammonium
20 mL sample; 1 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-20 Chloride
20 mL sample; 2 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-42 Fluoride
no correction possible.
1-44 Hydrazine
20 mL sample; 2 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-48
3
Silica
(high sensitivity)
20 mL sample; 1 mL sulfuric acid 10 %, mix; 1 mL R2, mix; 1mL R3, mix;
fill to 25 mL with silica-free water, mix.
1-51 Cobalt
20 mL sample; 1 mL R1, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-63 Nitrate Z
20 mL sample; 1 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-65 Nitrate
4.0 mL Nitrate R1, 0.5 mL sample, 0.5 mL 2-propanol, mix.
Blank value for correction:
4.0 mL Nitrate R1, 0.5 mL distilled water, 0.5 mL 2-propanol, mix.
1-67 Nitrite
20 mL sample; 2 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-75 Phenol
20 mL sample; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-77 Phosphate
20 mL sample; 1 mL sulfuric acid 30 %, mix; 1 mL R2, mix;
fill to 25 mL with distilled water, mix.
1-78 Phosphate
20 mL sample, 1.5 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-95 Zinc
20 mL sample; 1 mL R1, mix; 1 mL DMSO, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.