40a
500 D 01/2012
5.11.3 Determination of correction values for NANOCOLOR
®
standard tests with
sample blank value
For the tests of this category, the correction can already be performed during the
measurement. In these cases the following preparations are used as blank values
instead of the blank values given in the instructions.
For some tests, a special reagent has to be prepared for correction. If reduced
sample volumes are used for a test (dilution due to concentrations which are too
high), the sample volume for determination of the correction blank value must be
reduced as well.
Test
Preparation for correction blank value
1-16 free chlorine
total chlorine
20 mL sample; 1 mL R1, mix;
fill to 25 mL with distilled water, mix.
20 mL sample; 1 mL R1, mix; 5 drops of R3, mix;
fill to 25 mL with distilled water, mix.
1-25 Chromate
2,0 mL R2; 20 mL sample, mix;
fill to 25 mL with distilled water, mix.
1-30 Cyanide
20 mL sample; 1 spoon R1, dissolve; 2 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-36 Iron
20 mL sample; 1 mL R1, mix; 1 spoon R2, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-48 Silica
20 mL sample; 1 mL sulfuric acid 10 %, mix; 1 mL R2, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-53 Copper
20 mL sample; 2 mL R1, mix;
fill to 25 mL with distilled water, mix.
1-60 Manganese
20 mL sample; 1 mL R2, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
1-62 Nickel
20 mL sample; 1 mL R1; 1 mL R2, mix; 1 mL R3, mix;
fill to 25 mL with distilled water, mix.
Note: The yellow color produced by the addition of R2 must disappear when
R3 is added!
1-88 Sulfide
20 mL sample; 1 mL sulfuric acid 50 %, mix;
fill to 25 mL with distilled water, mix.