59
MD 610_1c 09/2015
13. Place the vial in the sample chamber making sure that
the
marks are aligned.
14. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
1. After use replace the bottle caps securely noting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C
.
3. Also see page 47
T1 accepted
prepare T2
press TEST
Countdown
2:00
*,** mg/l free Cl
*,** mg/l comb. Cl
*,** mg/l total Cl
Reagent
Form of reagent/Quantity
Order-No.
Set
DPD No. 1 buffer solution
DPD No. 1 reagent solution
DPD No. 3 solution
(approx. 300 tests)
3 x Liquid reagent / 15 ml
1 x Liquid reagent / 15 ml
2 x Liquid reagent / 15 ml
471056
DPD No. 1 buffer solution
Liquid reagent / 15 ml
471010
DPD No. 1 reagent solution
Liquid reagent / 15 ml
471020
DPD No. 3 solution
Liquid reagent / 15 ml
471030
Содержание MD 610
Страница 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Страница 2: ......
Страница 4: ......
Страница 16: ...8 MD 610_1c 09 2015 ...
Страница 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Страница 23: ...15 MD 610_1c 09 2015 ...
Страница 85: ...77 MD 610_1c 09 2015 ...
Страница 93: ...85 MD 610_1c 09 2015 ...
Страница 277: ...269 MD 610_1c 09 2015 Notes 1 SO3 Na2 SO3 Reagent Form of reagent Quantity Order No SULFITE LR Tablet 100 518020 ...
Страница 294: ...286 MD 610_1c 09 2015 ...
Страница 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Страница 298: ...290 MD 610_1c 09 2015 ...
Страница 350: ...342 MD 610_1c 09 2015 ...
Страница 351: ...343 MD 610_1c 09 2015 Part 3 Enclosure ...