60
MD 610_1c 09/2015
Chlorine, free
with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add the contents of
one VARIO Chlorine FREE-DPD /
F10 Powder Pack
straight from the foil to the water
sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the
marks are aligned.
8. Press
TEST
key.
The result is shown in the display in mg/l free
Chlorine.
Notes:
See page 47
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
1
2
3
4
0
1
1
Содержание MD 610
Страница 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Страница 2: ......
Страница 4: ......
Страница 16: ...8 MD 610_1c 09 2015 ...
Страница 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Страница 23: ...15 MD 610_1c 09 2015 ...
Страница 85: ...77 MD 610_1c 09 2015 ...
Страница 93: ...85 MD 610_1c 09 2015 ...
Страница 277: ...269 MD 610_1c 09 2015 Notes 1 SO3 Na2 SO3 Reagent Form of reagent Quantity Order No SULFITE LR Tablet 100 518020 ...
Страница 294: ...286 MD 610_1c 09 2015 ...
Страница 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Страница 298: ...290 MD 610_1c 09 2015 ...
Страница 350: ...342 MD 610_1c 09 2015 ...
Страница 351: ...343 MD 610_1c 09 2015 Part 3 Enclosure ...