155
MD 610_1c 09/2015
10. Press
ZERO
key.
11. Remove the vial from the sample chamber and empty
the vial.
12. Add
10 ml prepared water sample to the same
vial
.
13. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
10 drops KS61 (Ferrozine/Thioglycolate)
14. Close the vial tightly with the cap and swirl several times
to mix the contents.
15. Place the vial in the sample chamber making sure that
the marks are aligned.
16. Press
TEST
key.
Wait for a reaction
period of 5 minutes
(note 1,
page 153).
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total iron or, if
a filtered sample was used, in mg/l total soluble iron.
prepare Zero
press ZERO
Countdown
5:00
Zero accepted
prepare Test
press TEST