169
MD 610_1c 09/2015
Notes:
1. This test is applicable for the determination of soluble Manganese in water and
wastewater.
2. Highly buffered water samples or extreme pH values may exceed the buffering capacity
of the reagents and requires sample pre-treatment.
If samples were acidified for storing, adjust the pH between 4 and 5 with 5 mol/l (5 N)
Sodium hydroxide before test. Do not exceed pH 5, as manganese may precipitate.
3. Interferences:
Interfering substance
Interference level
Calcium
greater than 700 mg/l
Chloride
greater than 70 000 mg/l
Iron
greater than 5 mg/l
Magnesium
greater than 100 000 mg/l
4. Mn
MnO
4
KMnO
4
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Manganese Citrate Buffer F10
VARIO Sodiumperiodate F10
Powder Pack / 100
Powder Pack / 100
535100