140
MD 610_1c 09/2015
Iodine
with Tablet
0.05 – 3.6 mg/l I
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber,
empty the
vial leaving a view drops in.
5. Add
one DPD No. 1 tablet
straight from the foil to the
water sample and crush the tablet using a clean stirring
rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the
marks are aligned.
9. Press
TEST
key.
The result is shown in the display in mg/l Iodine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
1
5
2
Содержание MD 610
Страница 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Страница 2: ......
Страница 4: ......
Страница 16: ...8 MD 610_1c 09 2015 ...
Страница 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Страница 23: ...15 MD 610_1c 09 2015 ...
Страница 85: ...77 MD 610_1c 09 2015 ...
Страница 93: ...85 MD 610_1c 09 2015 ...
Страница 277: ...269 MD 610_1c 09 2015 Notes 1 SO3 Na2 SO3 Reagent Form of reagent Quantity Order No SULFITE LR Tablet 100 518020 ...
Страница 294: ...286 MD 610_1c 09 2015 ...
Страница 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Страница 298: ...290 MD 610_1c 09 2015 ...
Страница 350: ...342 MD 610_1c 09 2015 ...
Страница 351: ...343 MD 610_1c 09 2015 Part 3 Enclosure ...