321
MD 610_1c 09/2015
Fluoride Method 170
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [3] keys.
The display shows:
1. Fill a clean vial (24 mm Ø) with exactly
10 ml of deionised
water
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
exactly 2 ml SPADNS reagent solution
to the
water sample.
Caution: Vial is filled up to the top!
6. Close the vial tightly with the cap and swirl gently
several times to mix the contents.
<Calibration>
M170 Fluoride L
Zero: deionised water
press ZERO
<Calibration>
1: M191 Ca-Hardness 2
2: M191 0 Jus. Reset
3: M170 Fluoride L
Press [Shift] + [0] keys to keep the method blank.
Press [Shift] + [1] keys to erase the method blank and set
the value back to factory calibration.
The instrument goes back to mode menu automatically.
Mode
0
4
0
1
Shift
Shift
3
Shift
Содержание MD 610
Страница 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Страница 2: ......
Страница 4: ......
Страница 16: ...8 MD 610_1c 09 2015 ...
Страница 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Страница 23: ...15 MD 610_1c 09 2015 ...
Страница 85: ...77 MD 610_1c 09 2015 ...
Страница 93: ...85 MD 610_1c 09 2015 ...
Страница 277: ...269 MD 610_1c 09 2015 Notes 1 SO3 Na2 SO3 Reagent Form of reagent Quantity Order No SULFITE LR Tablet 100 518020 ...
Страница 294: ...286 MD 610_1c 09 2015 ...
Страница 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Страница 298: ...290 MD 610_1c 09 2015 ...
Страница 350: ...342 MD 610_1c 09 2015 ...
Страница 351: ...343 MD 610_1c 09 2015 Part 3 Enclosure ...