175
MD 610_1c 09/2015
13. Remove the vial from the sample chamber.
14. Place the sample in the sample chamber making sure
that the marks are aligned.
15. Press
TEST
key.
The result is shown in the display in mg/l Molybdate /
Molybdenum.
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 3 and pH 5 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionised water.
3. MoO
4
Mo
Na
2
MoO
4
Reagent / Accessories
Form of reagent/Quantity
Order-No.
Set
VARIO Molybdenum 1 LR F20
VARIO Molybdenum 2 LR
Powder Pack / 100
Liquid reagent / 50 ml
535450
Mixing Cylinder
25 ml
19802650
Zero accepted
prepare Test
press TEST