110
MD 610_1c 09/2015
Cyanide
with Reagent Test
0.01 – 0.5 mg/l CN
1. Fill a clean vial (24 mm Ø) with
2 ml of the water
sample and 8 ml of deionised water
, close tightly
with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
two level spoons No. 4 (white) of Cyanide-11
into the prepared water sample, replace the cap tightly
and invert the vial several times to mix the contents.
6.
Add two level spoons No. 4 (white) of Cyanide-12
,
replace the cap tightly and invert the vial several times
to mix the contents.
7. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
3 drops of Cyanide-13
8. Close the vial tightly with the cap and invert several
times to mix the contents.
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
Wait for a reaction
period of 10 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Cyanide.
Countdown
10:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
5
7
1
Содержание MD 610
Страница 1: ...Photometer 1 Instruction manual GB Lovibond Water Testing Tintometer Group www lovibond com ...
Страница 2: ......
Страница 4: ......
Страница 16: ...8 MD 610_1c 09 2015 ...
Страница 17: ...9 MD 610_1c 09 2015 Part 1 Methods ...
Страница 23: ...15 MD 610_1c 09 2015 ...
Страница 85: ...77 MD 610_1c 09 2015 ...
Страница 93: ...85 MD 610_1c 09 2015 ...
Страница 277: ...269 MD 610_1c 09 2015 Notes 1 SO3 Na2 SO3 Reagent Form of reagent Quantity Order No SULFITE LR Tablet 100 518020 ...
Страница 294: ...286 MD 610_1c 09 2015 ...
Страница 295: ...287 MD 610_1c 09 2015 Part 2 Instrument Manual ...
Страница 298: ...290 MD 610_1c 09 2015 ...
Страница 350: ...342 MD 610_1c 09 2015 ...
Страница 351: ...343 MD 610_1c 09 2015 Part 3 Enclosure ...