29
SETUP TESTING: “Settings” dropdown (Pg 2 of 2)
General Settings:
Emergency Start Pressure (PSI)
Service Start Pressure (PSI)
Emergency Air Loss Threshold (PSI)
Service Air Loss Threshold (PSI)
Air Leak Test:
Set your preferred values for testing. (Values originally set to logical defaults by Lite-Check.)
Test Length (seconds) Default = 60. Allowable range = 10 - 600 seconds.
Emergency Start Pressure (PSI) Default = 100. Allowable range = 65 - 105 PSI.
Service Start Pressure (PSI) Default = 100. Allowable range = 65 - 105 PSI.
Emergency Air Loss Threshold: (Default =2). Range: 1 - 5 PSI.
PSI loss at which the test is marked “FAIL”.
Service Air Loss Threshold: (Default =3). Range: 1 - 5 PSI.
PSI loss at which the test is marked “FAIL”.
121 Timing Test:
This test will perform the brake response time and display the results. Settings are adjustable
by using the blue slider on each of the three available items:
Input Pressure Max Threshold
Default = 102 PSI. (Allowed range: 101–-105 PSI.)
Emergency Pressure Timeout (Minutes)
Default = 5 minutes. (Allowed range: 1–-10 minutes.)
Service Release Build Pressure Timeout (Seconds)
Default = 10 seconds. (Allowed range: 5–-15 seconds.)
System Settings:
PDF Save Location
This setting will be a location for the test results to be stored. Once
selected, the location of the folder will be typed out next to the “Pick Location” selection box.
Lite-Check CERTIFIER 1200 Operations Manual
You must select “Unlock Manager Options”
to edit any values on the ‘Settings’ page.
Use the Scroll Bar to see this
lower part of the ‘Settings’ page.