55
15.1 Patient surface electric motors
Electrical connection
100-230 V AC 60 Hz
Protection type
IP 44
Protection class
II
Output voltage
30 V depending on model variant (see nameplate)
Operating time of motors
max. 2 min. adjustment operation, 18 min. pause
Power consumption
up to approx. 320 VA peak load
Adjustment speed
ca. 5,4 mm / sek.
Compression force before redirection
max. 6.000 N (Motor without emergency release)
max. 4.500 N ( Motor with emergency release )
15.2 Height adjustment electric motors
Max. voltage in idle mode
24 V
Current limiting
max. 10 A
Max. direct lifting force
2x 6.000 N
Max. tensile load
3.000 N
Adjustment speed
10-14 mm / sec. (depending on model, features, load)
Protection type
IP 44
Protection class
II Protective extra-low voltage
Operating time of motors
max. 2 min. adjustment operation, 18 min. pause