8.5 IV rod
The infusion stand is only intended for attaching infusions. Infusion pumps must not be
attached.
The maximum load for the infusion stand is 8 kg (17,64 lb). Each hook can be loaded
with 2 kg (4,41 lb).
The holder for the infusion stand is located at the head end of the patient surface frame.
If you want to attach an infusion stand:
• Insert the infusion stand from above into its holder on the inner side of the head end.
• Turn the infusion stand so that the bolt of the holder sleeve lies in one of the notches provided
in the lifting pole holder.
8.6 Patient surface extension
The sentida 03 offers two methods to extend the patient surface. A permanent extension of the
patient surface, carried out at the manufacturer´s plant, or an optional temporary extension.
Maximum 2 kg (4,41 lb) per hook
Infusion stand
Danger!
If you fail to close the foot-end gap resulting from the patient surface extension
with a protector and an mattress extension pad, there is danger of entrapment!
This applies for both the temporary and the permanent patient surface extension.
Alternatively a longer mattress can be used.