LightLink-CXL – Service Manual
Rev. No. : 01
Page
84
of
98
manufacturer for further assistance
Err 08 - Calibration
Error
-The system is
uncalibrated
-Call the authorized service or
manufacturer for further assistance
Err 09 – ROM Error
-System failed to pass the
software integrity test
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 10 – Watchdog
Error
-System watchdog flag
-Reboot the system
-Call the authorized service or
manufacturer for further assistance
Err 11 -
Communication
Error
-No communication
between optical driver
and the console within
the time of frame of
300ms during UV firing
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 12 - Power Error -The system power is
failure
-Call the authorized service or
manufacturer for further assistance