LightLink-CXL – Service Manual
Rev. No. : 01
Page
31
of
98
Figure 4.21 Treatment mode
Power Control and Laser Head
-
Power on by turning the key-switch and validate if the green LED is able to lit
-
Recheck if the aiming are visible or in place displaying the two dots
-
Validate if the LCD and laser head module established connection
System Initialization and Validation
Once everything validated then proceed to system initialization process
A.
Control Panel Integrity
-
Randomly select the treat mode
-
Validate if there is communication protocol establish
-
Validate all the controls are functioning accordingly such as, aiming intensity,
Backlight…etc.
-
Execute the UV process
-
B.
Aiming Integrity Validation
-
Engage the aiming intensity to see if the aiming spots are evenly distributed
-
Proceed to calibration process if not valid (refer to section 7 on page 72)