D. Spot Size Power Calibration (Power Spot Parameter)
This procedure is intended to calibrate the individual spot sizes power for the system
spot power consistency and accuracy during the treatment procedure.
- Enter to calibration mode
- Press the "Power Spot Parameter" icon to initial the calibration process
- Set the power meter wavelength at 365nm and set the spot size to 11mm in main
calibration menu (refer to motor parameter) and power to 30 mW
- Start UV light run icon and measure if the value falls within the calibration
specification guideline(Refer to below figure)
Figure 7.16 Power Spot Parameter