Section 7 Calibration / Alignment Procedures
7.1
Calibration Procedures
Purpose
The purpose of this procedure is to calibrate or restore the laser system after any
module replacement or during any routine checkout process. Prior to commencing the
procedure, only the qualified or trained personnel is permitted to conduct this process.
Performing these procedures will reset the all the factors of the laser system.
The procedures can be defined into four parts:
- Motor Aperture Calibration (Motor Parameter)
- Power Calibration (Power Parameter)
- Detector Calibration (Detector Parameter)
- Spot Size Power Calibration (Power Spot Parameter)
Equipments / Tools required:
Interlock Connector
Calibrated Laser Power Meter
Safety Goggle (for this specified system)
VR pot adjuster or small flat screw driver
Note: Do remember to remove the front cover before any realignment process
takes place