LightLink-CXL – Service Manual
Rev. No. : 01
Page
57
of
98
A. Motor Aperture Calibration (Motor Parameter)
This procedure will calibrate the motor aperture position which allows the system to
have the corrected spot size during the any treatment procedure. Improper spot size
will affect the treatment quality.
- Access to calibration mode
- Double clicks the hidden icon at the left top corner screen (refer to fig 7.1)
- Enter the pass code "0000"
Figure 7.1 Pass code entering
- Restart the main program again by click the software program in startup menu
Figure 7.2 The software position in Main Program
Hidden Icon
Pass code: '0000'
Double clicks to initial
the software