Command
Register
The Command Register (40002) is used to make the device perform an
action. The register performs an action when an integer value is written
to it. The action is completed when the device sends a MODBUS
1
ApexRemote:
LASER CURRENT STATUS:
Set to 1 when unit’s
LASER power is out of spec, else set to 0
2
ApexRemote:
LASER SUPPLY STATUS:
Set to 1 when unit’s
LASER supply is out of spec, else set to 0
3
ApexRemote:
LASER LIFE STATUS:
Set to 1 when unit’s LASER
supply is out of spec, else set to 0
4
ApexRemote:
NO FLOW STATUS:
Set to 1 when unit’s flow is below
no flow threshold causing LASER to turn off, else set to 0
5
ApexRemote:
PHOTOAMP SUPPLY STATUS:
Set to 1 when unit’s
photoamp supply is out of spec, else set to 0
6
ApexRemote:
BACKGROUND STATUS:
Set to 1 when unit’s pho-
toamp background is out of spec, else set to 0
7
ApexRemote:
PHOTODIODE STATUS:
Set to 1 when photodiode
has failed, else set to 0
8
ApexRemote:
CALIBRATION DUE DATE STATUS:
Set to 1 when
unit is past calibration due date, else set to 0
9
ApexRemote:
LOCATION BRACKET STATUS:
Set to 1 when unit
is in location bracket mode and bracket is missing, else
set to 0
Table A-6 Device Status Status (40056)
Bit
Description
Содержание ApexR02
Страница 1: ......
Страница 2: ...00 ...
Страница 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Страница 6: ......
Страница 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Страница 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Страница 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Страница 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Страница 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Страница 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Страница 83: ...00 ...
Страница 84: ......