Communications
248083447-1 Rev 1
4-5
RS-485 Communications
RS-485 must be used if the instrument is more than 50 feet from a
computer or is installed in an industrial network. Refer to Table 4-5 for
specifics about RS-485. Contact Lighthouse Technical Support for
more information.
Table 4-5 shows the Electronics Industry Association (EIA) industry
standards RS485 specifications.
Table 4-5 EIA Industry Standards for RS-485 Communications
SPECIFICATIONS
RS-485
Mode of Operation
Differential
Total Number of Drivers and Receivers on
One Line (One driver active at a time for
RS-485 networks)
32 Drivers
32 Receivers
Maximum Cable Length
4000 ft (1,219.2 m)
Maximum Data Rate (40 ft - 4000 ft for
RS422/RS-485)
100Kbs - 10Mbs
Maximum Driver Output Voltage
-7V to +12V
Driver Output Signal Level (Loaded
Min.): LOADED
+/-1.5V
Driver Output Signal Level (Loaded
Max.): UNLOADED
+/-6V
Driver Load Impedance (Ohms)
54
Max Driver Current in High Z State
(POWER ON)
+/-100
μ
A
Max Driver Current in High Z State
(POWER OFF)
+/-100
μ
A
Receiver Input Voltage Range
-7V to +12V
Receiver Input Sensitivity
+/-200mV
Receiver Input Resistance (Ohms), (1
Standard Load for RS-485)
>12k
Содержание ApexR02
Страница 1: ......
Страница 2: ...00 ...
Страница 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Страница 6: ......
Страница 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Страница 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Страница 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Страница 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Страница 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Страница 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Страница 83: ...00 ...
Страница 84: ......