
Lighthouse ApexRemote Operating Manual
4-4
248083447-1 Rev 1
Figure 4-2 Serial COM Port
The connector pinouts are shown in Table 4-4.
To connect the instrument to an RS-485 network:
1. Make sure the
SmartPort
cable is disconnected from the
instrument.
2. Install the
Apex
Remote
in a perpendicular position with its Inlet
barb upward. Connect one end of a CAT5e cable to the Serial COM
port on the instrument (shown in Figure 4-2).
3. Connect the other end of the cable to an available RS485 port on an
LWS 485 Gateway
, an
LWS System Control Cabinet RS485
port or
other similar equipment port.
Table 4-4 RJ45 Pinouts
RJ45 Pin
Signal Name
1
RS-232-TX (Output)
2
RS-232 RX (Input)
3
RESERVED for future use
4
RS-485B
5
RS-485A
6
RESERVED for future use
7
24VDC
8
GROUND
Switch 1 through 5
Содержание ApexR02
Страница 1: ......
Страница 2: ...00 ...
Страница 3: ...Lighthouse Worldwide Solutions ApexRemote Airborne Particle Counter Operating Manual ...
Страница 6: ......
Страница 10: ...Lighthouse ApexRemote Operating Manual t iv 248083447 1 Rev 1 ...
Страница 12: ...Lighthouse ApexRemote Operating Manual ii 248083447 1 Rev 1 ...
Страница 22: ...Lighthouse ApexRemote Operating Manual 2 8 248083447 1 Rev 1 ...
Страница 32: ...Lighthouse ApexRemote Operating Manual 3 10 248083447 1 Rev 1 ...
Страница 76: ...Lighthouse ApexRemote Operating Manual A 18 248083447 1 Rev 1 ...
Страница 78: ...Lighthouse ApexRemote Operating Manual B 2 248083447 1 Rev 1 ...
Страница 83: ...00 ...
Страница 84: ......