LI-600 Porometer/Fluorometer Instruction Manual
8-18
serial number, the LI-600 will continue logging data
using the old sensor serial number.
Take the following steps to change the RH sensors serial
number.
1
Open Factory Calibration.
Connect the LI-600 to a computer via USB then open
the computer software. In Explore the LI-600, select
the drop-down menu on the Toolbar then select
About Instrument. Check the Factory Calibration
box then click
Close
. Select the drop-down menu
again to allow the software to register the change
then click
Factory Calibration
.
2
Calibrate the RH sensor.
Select RH under Factory Calibration then click
Cal-
ibrate
. Click
OK
on the confirmation prompt.
3
Change the RH sensor serial number.
The
Change RH Sensor Serial Number
and
Auto-
matic RH Sensor Calibration
buttons will appear.
Click
Change RH Sensor Serial Number
then
input the serial number. Click
OK
on the con-
firmation prompt.
4
Calibrate the RH sensor.
Click
Automatic RH Sensor Calibration
to cal-
ibrate the RH sensor.