LI-600 Porometer/Fluorometer Instruction Manual
8-14
1
Pump for drying the RH sensors.
Proper drying of the RH sensors can take up to 30
minutes, as water adheres to all surfaces in the flow
path. This ensures that no water remains in the sys-
tem.
Dry the sensors using one of two methods: using
user-supplied airstream or using the LI-600 pump. If
drying the sensors by delivering an airstream across
the sensors using either a tank of dry gas (recom-
mended) or an alternate airstream method, see
on page 8-11. If delivering air-
flow across the sensors via the LI-600 pump only, see
on page 8-8. If the resulting
RH is <1% and trending down after several minutes,
the zero does not require resetting.
2
Open Instrument Calibration.
Connect the LI-600 to a computer via USB then open
the computer software. Click the drop-down menu
on the Toolbar then select Instrument Calibration.
3
Power off the pump and power on the blower.
Select RH Zero then click
Begin RH Zero Cal-
ibration
to view the RH Ref and RH Sample. Click
TURN PUMP OFF
then click
TURN BLOWER ON
.
4
Zero the RH sensors.
When the RH stabilizes, click
Zero
for RH Ref then
again on the confirmation prompt to reset the zero.
Click
Apply
on the second confirmation prompt to
finish. Repeat for RH Sample.
RH sensor span
A known H
2
O input is required to span the RH sensors
using another H
2
O analyzer or a dew point generator.