![LI-COR LI-600 Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/li-cor/li-600/li-600_instruction-manual_1906650024.webp)
Section 3. Taking Measurements
3-3
l
Maintain unblocked vents on the instrument.
Do not cover any of the vents on the instrument
when taking measurements.
l
Turn GPS off when not in use.
Turning off
GPS will save battery power, especially in buildings
with poor satellite fix.
Caution:
Do not apply pressure to the aperture
clamp to achieve a seal on the leaf. This can warp the
aperture clamp, interfere with the instrument’s
optics, and produce measurement errors.
Taking a measurement
The workflow for taking a measurement on the LI-600
will vary depending on two settings in the active con-
figuration: whether the measurements are taken in
Auto mode
or
Manual mode
, and whether
Prompt
on log
is enabled for optional data labels.
Taking a measurement in Auto Mode
In
Auto Mode
, the LI-600 automatically logs a meas-
urement when the stability criteria set in the con-
figuration are met.
To take a measurement in Auto Mode:
1
Power on
the LI-600.
2
Select a configuration
with Auto mode enabled and
press Enter. The instrument will match and go to the Meas-
urement screen. If
Prompt on log
is not enabled and you
want to add data labels or scan a barcode, press the right
arrow (Prompt) to Data labels (
Press enter to edit high-
lighted values
). Select a data label, press the Enter but-
ton, and choose from the options. If you've enabled the