LI-600 Porometer/Fluorometer Instruction Manual
8-8
Calibrating the sensors
The LI-600 includes sensors whose calibration para-
meters can be adjusted by the user: relative humidity
(RH) sensors, flow sensors, and a fluorescence detector.
This section describes the process of setting the available
parameters for each sensor.
To view user calibration coefficients and change the
user-adjustable parameters, connect the LI-600 to the
computer software. In the drop-down menu on the Tool-
bar, select Instrument Calibration. (For sensor equa-
tions, see
All calibration coefficients can be viewed in
Factory
Calibration
if enabled in the software Preferences. See
RH sensors
RH sensor parameters include Zero, or the sensor’s
response to dry air, and Span, or the sensor’s response to
a known dew point or H
2
O mole fraction.
Using the LI-600 pump
Take the following steps to dry the RH sensors using the
internal pump and zero kit.
1
Prepare the LI-600.
Power off the instrument.
2
Remove the shells.
Remove the shells on the instrument head. Set the
shells safely aside (see
3
Remove the filter cover and filter.
Содержание LI-600