Section 8. Maintenance
8-15
Take the following steps to span the RH sensors.
1
Prepare the LI-600.
Power on the instrument.
2
Open Instrument Calibration.
Connect the LI-600 to a computer via USB then open
the computer software. Click the drop-down menu
on the Toolbar then select Instrument Calibration.
3
Set the H
2
O input.
Select RH Span then click
Begin RH Span Cal-
ibration
to view the RH Ref and RH Sample. Select
either Dew Point or Mole Fraction then enter the
desired H
2
O input.
4
Power off the pump.
Select Utility then click
TURN PUMP OFF
.
5
Remove the leaf clamp.
Gently but firmly pull on the leaf clamp to remove it
from the instrument.
Figure 8-13
. Leaf clamps.
6
Attach the calibration clamp.