LI-600 Porometer/Fluorometer Instruction Manual
5-8
Figure 5-6
. The Measurement Settings screen.
If the
Dark Adapted
box is checked or the
Rect-
angular
or
Multiphase Flash
options are selected, val-
ues in Flash and Modulation automatically set to
default.
High flow
is recommended for most applications to
provide the fastest measurement speed with no observed
reduction in sensitivity.
Match frequency
determines how often the instru-
ment will automatically match. Enter the interval in
minutes.
Flash
(fluorometer) sets the flash to light (default) or
dark adapted, and whether the flash will be rectangular