LI-600 Porometer/Fluorometer Instruction Manual
8-10
Figure 8-8
. The buffer material.
6
Reinstall the filter cover, filter, pump cover, and pump.
Reinstall the parts with the H
2
O buffer material
removed.
7
Connect the zero kit.
Attach the zero kit to the air inlet next to the tripod
mount via the 10-32 fitting.
Figure 8-9
. Attach the zero kit.
8
Seal the leaf aperture.
To seal the leaf aperture, clamp on a plastic bag or
similar material.
9
Open Instrument Calibration.
Power on the LI-600. Connect it to a computer via
USB then open the computer software. Click the