LI-600 Porometer/Fluorometer Instruction Manual
7-6
configuration; the blower should automatically turn on.
If the warning persists, power on the LI-600 then con-
nect it to a computer via USB. Open the computer soft-
ware. From the drop down menu select Instrument
Calibration. Select and begin any of the four cal-
ibrations then
Turn Blower On
.
Pump
The Pump warning appears when the Flow In sensor is
< 90% of the flow set point (i.e., unable to reach the
desired flow). Ensure that the inlet is not blocked. If the
warning persists, ensure that the filter is clean; if it is
dirty, replace the filter. If the warning persists, replace
the pump (see
No GPS map pin
If no GPS map pin appears in the top of the instrument
display, go to Settings and make sure GPS is enabled. If
enabled and still no map pin, go to GPS settings from
Instrument Settings
on the device and check
NumSats
; if <3, move to a location away from buildings
and wait to see if GPS can get a satellite lock. If you are
working inside, you may not have the necessary recep-
tion. GPS can be turned off to conserve battery power.
Data download errors
Data download error messages pop up with "Data
Download Failed!" However, if a data download error
occurs, the raw data file is still saved to your local com-
puter. Find and open the data file, and reference the
error message to identify the problem. Following are
known data download errors and their causes.