LI-600 Porometer/Fluorometer Instruction Manual
5-6
Number
: Check the Auto-Increment box for automatic
numbering.
List
: Add up to three selections with the plus button;
drag and drop them to change their order.
Barcode
: Enable the barcode scanner.
Automode Setup
In
Automode Setup
, select Automode,
gsw
, and/or
F
(fluorometer).
Select Automode to access the
gsw
and
F
variable
options. At least one variable must be selected to enable
Auto Mode.
Figure 5-4
. Automode Setup screen.
Then select the variables (
gsw, F
, or both), then a Preset
(Fast, Medium, or Custom). If selecting Fast or
Medium: the Stability Limit (
gsw-
specific option), Slope
Limit (F-specific option), and Period over which to cal-
culate the change are automatically determined. If select-
ing Custom, you can set Stability Limit, Slope Limit,
and Period (1, 2, or 4 seconds). (See
Preset
Stability Limit
(g
sw
)
Slope Limit
(F)
Period
(s)
Fast
0.005
5
2
Table 5-1
. Automode Setup variable options.