Section 8. Maintenance
8-11
drop-down menu on the Toolbar then select Instru-
ment Calibration.
10
Power on the internal pump.
Select RH Zero then click
Begin RH Zero Cal-
ibration
to view the RH Ref and RH Sample. Click
TURN PUMP ON
.
11
Dry the RH sensors.
Wait for RH to stabilize at ±1% then select Zero in
the computer software. If trending down, zeroing is
not required.
Ensure that the chemicals are fresh, such as Drierite®
(P/N 622-10509), to provide a low dew point.
12
Reinstall all parts in reverse order.
Reinstall the H
2
O buffer material, pump, pump
cover, filter, filter cover, and instrument head shells.
Using a user-supplied airstream
Follow these steps when drying the RH sensors using a
user-supplied airstream from an air tank. Alternatively,
you can use a user-supplied pump and the zero kit (P/N
9968-258) with chemicals to dry the RH sensors. If using
an external pump with the zero kit, ensure that the
chemicals are fresh, such as Drierite, to provide a low
dew point.
Using an airstream in the RH sensor zeroing process
requires the least disassembly.
1
Prepare the LI-600.
Power on the instrument.
2
Open Instrument Calibration.