Calibration
GA10.218_1.01 - 11/2004
⏐
85
7.3 Factor of Calibration - Range
of Values
To avoid a faulty calibration the factor of calibration is tested for plausibility
at the end of the calibration routine:
When the new factor of calibration is not considerably higher or lower
(< factor 2) than the previous factor of calibration the new factor will be
accepted automatically. When the new factor of calibration diverges stron-
ger from the previous factor the user can decide if he wants to accept it
anyway (e. g. after changing the system configuration) or not (e. g.
because of a maloperation).
When calibration is started via SPS or RS232 no testing for plausibility is
occuring.
When calibrating internal it is also monitored if the newly calculated factor
of calibration is higher than 10 or lower than 0.1. In this case a warning
(see W81 resp. W82 in chapter
) is displayed and the calibration will be
interrupted.
• The bargraph display
shows a signal which
must not decrease
any more. There might
be a small fluctuation
which is okay. If so
please press OK (Soft
Key no. 8).
Fig. 33 External Calibration, Step 5
• When signal is stable
affirm with o.k.
Fig. 34 External Calibration, Step 6
• The Phoenix shows the
old and new calcula-
ted calibration factor.
Fig. 35 External Calibration, Step 7
Содержание PhoeniXL300
Страница 18: ...General Information 18 GA10 218_1 01 11 2004 ...
Страница 30: ...Operation Check 30 GA10 218_1 01 11 2004 ...
Страница 44: ...Working Principle 44 GA10 218_1 01 11 2004 ...
Страница 48: ...Operation 48 GA10 218_1 01 11 2004 refer to Chapter 6 2 1 Scale linear logarithmic Fig 15 Display Trend Mode ...
Страница 82: ...Menu 82 GA10 218_1 01 11 2004 ...
Страница 86: ...Calibration 86 GA10 218_1 01 11 2004 ...
Страница 98: ...Maintenance 98 GA10 218_1 01 11 2004 ...
Страница 101: ......