GA10.218_1.01 - 11/2004
⏐
53
6.2.2 Display-range auto/manual
>
>
The upper limit of the displayed leak rate range can be set manually or
automatically. These settings apply to the bargraph ( = bar underneath the
digital figures in the measurement mode and y-axis in the trend mode).
With the upper limit defined here the lower limit is set to a value based on
the number of decades (See Chapter
).
If linear scale is selected, the lower limit is always zero. The upper limit is
only a default value. You can change this on the measurement screen with
the Soft Key 6 and 7 if you have chosen manual display ranging.
6.2.3 Time axis
>
>
The length of the time axis in trend mode can be changed in given steps
between 16 … 960 seconds.
Softkey 2: Manual
The upper limit of the displayed range can be set manually.
Softkey 3: Arrow down
Decrease the upper limit if manual is chosen. The minimum value is
10
-11
mbar l/s
Softkey 5: ?
Helpfunction
Softkey 6: Automatic
The limit of the displayed range will be chosen automatically.
Softkey 7: Arrow up
Increase the upper limit if manual is chosen. The maximum value is
10
+3
mbar l/s
Softkey 8:
Save the settings and return to the previous menu.
Softkey 3: Arrow down
Decrease the length of the time axis. The minimum value is 16 seconds.
Softkey 5: ?
Helpfunction
Softkey 7: Arrow up
Increase the length of the time axis. The maximum adjustable value is 960
seconds.
Содержание PhoeniXL300
Страница 18: ...General Information 18 GA10 218_1 01 11 2004 ...
Страница 30: ...Operation Check 30 GA10 218_1 01 11 2004 ...
Страница 44: ...Working Principle 44 GA10 218_1 01 11 2004 ...
Страница 48: ...Operation 48 GA10 218_1 01 11 2004 refer to Chapter 6 2 1 Scale linear logarithmic Fig 15 Display Trend Mode ...
Страница 82: ...Menu 82 GA10 218_1 01 11 2004 ...
Страница 86: ...Calibration 86 GA10 218_1 01 11 2004 ...
Страница 98: ...Maintenance 98 GA10 218_1 01 11 2004 ...
Страница 101: ......