Maintenance
GA10.218_1.01 - 11/2004
⏐
95
tools. (Cat. no. 20099025)
Filter mat c) This filtermat is fixed before the fan with a plastic bracket.
Take of the bracket and remove the filter, put on the bracket and fix it pro-
perly. ( cat. no. 200001366)
Under certain circumstances a dirty mat may be cleaned by shaking the
dust out or by using a vacuum cleaner so that the filter mat be reused.
9.1.4 Exchanging the oil
Remove the cover of the mechanical section as described in Chapter
9.1.2.
Caution
Before starting with any disassembly work on the pumps, pull the
mains plug first. When the pump has been pumping hazardous
substances, dertermind the kind of hazard first and ensure that
suitable safety precautions are taken.
Observe all safety regulations
Attention
When disposing of waste oil you must observe the applicable
regulations for the safety of the environment.
The oil change procedure are described in the corresponding Operating
Instructions GA 01.601 and these must be foolowed closely.
As already stated in Chapter 9.1.1 only Arctic oil must be used for the D
2,5E pump in the Phoenix.
9.1.5 Cleaning
The housing of the Phoenix is made of painted plastic parts. Thus for the
purpose of cleaning, only such agents should be used which are generally
also used for other painted or plasctic surfaces (mild household cleaning
agents, for example). Normally a moistened peace of cloth will do. Never
use any solvents which are capable of dissolving paint (like acetone, toluol,
etc.).
A soft brush or a vacuum cleaner is recommended for cleaning the ventila-
tion slits.
9.1.6 Exchanging the fuses
Caution
Before exchanging the fuses you must disconnect the mains
cord.
1. Switch the Phoenix off.
2. Pull the mains cord off the Phoenix.
3. Use a screwdriver to fold out the lid of the mains socket from the right
(the mains switch is not affected by this).
4. The fuses can be removed by pulling the drawers out which are marked
by the arrows. When reinsering these make sur that the arrows point
Содержание PhoeniXL300
Страница 18: ...General Information 18 GA10 218_1 01 11 2004 ...
Страница 30: ...Operation Check 30 GA10 218_1 01 11 2004 ...
Страница 44: ...Working Principle 44 GA10 218_1 01 11 2004 ...
Страница 48: ...Operation 48 GA10 218_1 01 11 2004 refer to Chapter 6 2 1 Scale linear logarithmic Fig 15 Display Trend Mode ...
Страница 82: ...Menu 82 GA10 218_1 01 11 2004 ...
Страница 86: ...Calibration 86 GA10 218_1 01 11 2004 ...
Страница 98: ...Maintenance 98 GA10 218_1 01 11 2004 ...
Страница 101: ......