Calibration
84
⏐
GA10.218_1.01 - 11/2004
messages are displayed and the described actions are required:
• Make sure that the test
leak is connected and
opened.
• Check the leak rate
printed on the test
leak and compare it
with the leak rate at
the display. If the leak
rates are not identical
press Edit leak rate
(Soft Key no. 4) and
correct the value.
• If the leak rates are
okay press START
(Soft Key no. 8).
Fig. 29 External Calibration, Step 1
• No action required.
Fig. 30 External Calibration, Step 2
• The bargraph display
shows a signal which
must not vary much. If
so please press OK
(Soft Key no. 8).
Fig. 31 External Calibration, Step 3
• No action required.
Fig. 32 External Calibration, Step 4