![LeCroy WaveRunner 6Zi Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372087.webp)
WaveRunner 6 Zi Oscilloscopes
77
WR6Zi-GSM-E
The trigger location is shown by a marker under the grid.
Post-trigger delay is indicated by a left-pointing arrow at the
lower-left of the grid.
The time value is given in the title line of the
Timebase
label at the
lower-right of the grid.
Vertical:
Turn the
L
EVEL
knob in the TRIGGER control group to
adjust the trigger's vertical threshold.
Turn this knob to adjust the level of the trigger source or the
highlighted trace. Level defines the source voltage at which the
trigger will generate an event: a change in the input signal that
satisfies the trigger conditions.
Alternatively, in the Trigger dialog, you can touch inside the
Level
control and provide a value using your preferred input control
method - for more information regarding the touch screen
controls, refer to
Dialog Area
(on page 46). Quickly set a level of
zero volts by touching the
Zero Level
button.
An arrow on the right side of the grid shows the threshold
position. This arrow is only visible if the trigger source is displayed.
Trigger Setup
Based on your
Trigger Type
selection, sections of the main Trigger dialog
and additional dialogs altogether vary.
Most trigger types have
Type
,
Setup
, and
Level
selections. The following
sections explain some of the standard setup configurations for different
trigger types.